Features
✧ The uniquely designed optical system, high-performance holographic grating and receiver ensure that the instrument has excellent performance indicators;
✧ Cooperate with PC anti-control workstation, the function and data processing capabilities are very powerful;
✧ Use the instrument itself to operate linear regression equation measurements, etc.;
✧ Software applying the latest processing technology to make operation more convenient.
Applications:
Photometric measurements;
Quantitative measurement;
Qualitative measurement (operated on PC);
Kinetic analysis (operated on PC).
Parameters:
Model No. | T-6 | D-7 |
Slit width | 2nm | 2nm |
Light system | Single beam | Single beam |
Wavelength range | 190~1100nm (190-339.9nm UV area; 340-1100nm visible area) | 190~1100nm (190-339.9nm UV area; 340-1100nm visible area) |
Wavelength accuracy | ±1nm | ±0.5nm |
Wavelength repetition rate | ±0.2nm | ±0.2nm |
Wavelength resolution | 0.1nm | 0.1nm |
Photometric range | -0.3~3ABS | -0.3~3ABS |
Photometric accuracy | ±0.5%T (0~100%T) | ±0.3%T (0~100%T) |
Photometric Repeatability | ±0.2%T (0~100%T) | ±0.1%T (0~100%T) |
Stray light | ≤0.2%T | ≤0.1%T |
Baseline straightness | / | ±0.002A (190~1100nm) (operate on PC) |
Stability | ±0.002A/hr@500nm, 0A | ±0.002A/hr@500nm, 0A |
Photometric noise | / | ±0.002A |
Display | 128×64 graphic display LCD screen | 128×64 graphic display LCD screen |
Data output | Connect to upper-level software output | Connect to upper-level software output |
PC anti-control software | Optiional | Optional |
Wavelength scan | Available with optional software | Available with optional software |
Outer dimension | 430*330*200mm | 470*400*230mm |
Net weight | 10kg | 16kg |
Accuracy level | Level 3 | Level 2 |